Patient Orientation Modifier is not extracted from RDSR
Issue #561
resolved
Patient Orientation Modifier is missed because we're looking for it at the same nesting level as Patient Orientation, but it is one nesting level deeper (as it is a concept modifier of Patient Orientation).
Comments (4)
-
reporter -
Adding ref
#561to changes→ <<cset dac898630430>>
-
- changed status to resolved
Merged in Issue561PatientOrientationModifier (pull request #129)
Fixes
#561Extract Patient Orientation Modifier correctlyTested import using file previously provided. @LuukO it would be good to have version with just a few events in to add to the automated testing. Adding this one would make the tests take too long!
→ <<cset 8ea8fd6b32a4>>
-
reporter @edmcdonagh you can remove this test file and use RF-RDSR-Philips_Allura.dcm for the purpose. It is the same concerning this issue.
- Log in to comment
Refs
#561Extract Patient Orientation Modifier correctlyMade the extraction one nesting level deeper (as it is concept modifier of Patient Orientation)
→ <<cset 3086b2d75998>>