Fluoro exports report total time as ms - should be s

Issue #573 resolved
Ed McDonagh created an issue

Accumulated times in RDSRs are in seconds, event times are in milliseconds.

Export headers include Total fluoro time (ms) and total acquisition time (ms) when in fact both values are in s.

Issue highlighted by Kirsten Nygaard Bolstad in this post to the google group

Comments (3)

  1. Ed McDonagh reporter

    Corrected the flouro exports to show summary total times in seconds, ref #573. This refers to irradiation duration (including gaps between pulses), not exposure time (x-ray on time only). Added number of pulses and exposure duration to the series data in XLSX export. Moved pulse rate column along.

    → <<cset 3baf3707d691>>