Extractor for old Toshiba systems fails

Issue #617 resolved
David Platten created an issue

If an examination carried out on an older Toshiba system contains multiple studies then the extractor creates a sub-folder for each, and copies the corresponding images into each sub-folder. The extractor then creates an RDSR in each sub-folder, and then supplements this with data from image tags.

There is an assumption that each sub-folder contains a dose image file with suitable data to create the initial RDSR using the Pixelmed.jar utility. This assumption is incorrect, and the extractor fails if a sub-folder does not contain a dose image file.

Comments (8)

  1. David Platten reporter

    Fixed issue where examination with multiple studies caused the extractor to crash if one of the studies does not contain a dose summary file. References issue #617

    → <<cset 9eeb8522ba3b>>