- changed milestone to 0.9.2
GE OEC Elite Mini View RDSR has no Content Template Sequence
Issue #746
resolved
Which means the extraction routine never launches into the RDSR properly as there is no trigger to indicate what we should find.
Propose checking if nether the CT or projection templates have been found, and then attempt to work it out from there using Procedure Reported
etc.
Not sure of the milestone yet.
Issue raised by @hrkhosravi
Comments (8)
-
reporter -
reporter Fix for GE OEC mini with no template ID sequence. Refs
#746→ <<cset 0d7c73c8532e>>
-
reporter Added test for ref
#746→ <<cset aa90ca39faae>>
-
reporter Forgot to un-disable test! Refs
#746→ <<cset 6c504910d649>>
-
reporter Would help if I included the new DICOM file...! Refs
#746→ <<cset 72179717e9fc>>
-
reporter Adding ref
#746to change - still targeting 0.9.2→ <<cset e5564af27999>>
-
reporter - changed status to resolved
-
reporter - changed milestone to 0.10.0
- Log in to comment