Toshiba bi-plane fluoro crashes system

Issue #362 resolved
David Platten
created an issue

RDSR imports, but when trying to view in the filtered or detailed view the system crashes.

This is because there are two accumxraydose_set objects, rather than the usual one.

Comments (15)

  1. Ed McDonagh

    I don't suppose you have a QA example of this you could send me @David Platten?

    @john loveland - could this be the behaviour you have seen? I hadn't thought about it at the time you told me of your issue, but I now realise the cath lab systems I have in my database are all single plane - the biplane systems don't have RDSRs.

  2. Ed McDonagh

    Added in acquisition plane to RF listing, but only if there are two planes. Fixes #362. I've reviewed the import code, and I can't see any issues with the import of bi-plane fluoro, just the display which this issue addresses. Also removed an incidental link to 'list' which didn't work.

    → <<cset 6aae614693b5>>

  3. Ed McDonagh

    Added in acquisition plane to RF listing, but only if there are two planes. Fixes #362. I've reviewed the import code, and I can't see any issues with the import of bi-plane fluoro, just the display which this issue addresses. Also removed an incidental link to 'list' which didn't work.

    → <<cset 6aae614693b5>>