Issue #362 resolved

Toshiba bi-plane fluoro crashes system

David Platten
created an issue

RDSR imports, but when trying to view in the filtered or detailed view the system crashes.

This is because there are two accumxraydose_set objects, rather than the usual one.

Comments (15)

  1. Ed McDonagh

    I don't suppose you have a QA example of this you could send me David Platten?

    john loveland - could this be the behaviour you have seen? I hadn't thought about it at the time you told me of your issue, but I now realise the cath lab systems I have in my database are all single plane - the biplane systems don't have RDSRs.

  2. Ed McDonagh

    Added in acquisition plane to RF listing, but only if there are two planes. Fixes #362. I've reviewed the import code, and I can't see any issues with the import of bi-plane fluoro, just the display which this issue addresses. Also removed an incidental link to 'list' which didn't work.

    → <<cset 6aae614693b5>>

  3. Ed McDonagh

    Added in acquisition plane to RF listing, but only if there are two planes. Fixes #362. I've reviewed the import code, and I can't see any issues with the import of bi-plane fluoro, just the display which this issue addresses. Also removed an incidental link to 'list' which didn't work.