Fluoro pulse width, kV, current and exposure can all be multi-value...

Issue #584 resolved
Ed McDonagh created an issue

In http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_TID_10003B.html pulse width, kV, current and exposure can all be multi-valued, ie one value of each per pulse in that event.

The database is set up for this, but the extractor routine, display and export routines are not.

I think that at least the import routine needs to be fixed asap.

Comments (17)

  1. Ed McDonagh reporter

    Exposure/mAs displays and exports correctly. Average/mean mAs in export, list in interface. Modified uAs to mAs conversion to return if 0. Refs #584

    → <<cset f0edc876bd7b>>

  2. Ed McDonagh reporter
    • changed status to open

    Re-opening in light of discussions around issue ref #602. Will now remove zeros before calculating mean. Will then need to document in ref #586.

  3. David Platten

    Should be easy to do this - looks like a few exclude clauses need to be added. Do you want me to do it?