EOS Imaging unit is bi-plane DX and causes 500 errors in DX interface

Issue #722 resolved
Ed McDonagh
created an issue

DX filtered assumes single DTID 10002 “Accumulated X-Ray Dose”, but EOS system is a biplane system and therefore has two, as per bi-plane fluoro systems.

Issue raised by @Hamid Khosravi via email

Device in question has website at https://www.eos-imaging.us/us/professionals/eos/eos including links to DICOM conformance statement and IHE integration statement.

Comments (11)

  1. Ed McDonagh reporter

    I've checked in the RDSR I have been provided, and it seems that although the system lists plane A and plane B, the summary information for plane B is a dose of 0, and there is no exposure level information for plane B. This is despite, as far as I can make out, the RDSR being for a dual plane image.