Radiographic PHE 2019 Survey export
Radiography “PHE 2019 Survey: study” export option gives exam DAP in exported data that is 1,000,000 times smaller than sum of individual projection DAP values: i.e units of Gym² instead of intended cGycm². This need correcting.
Documentation states that export will be formatted for skeletal survey spreadsheet template if the number of exposures exceeds eight: this is in fact true for number of exposures exceeding six. Documentation needs correcting.
Exports from our data have units of DAP as “cGycm²” which doesn’t match drop-down option in PHE spreadsheet of “cGy x cm2 (or shown as CGYCM2)”. I don’t know if this is important or not.
Comments (4)
-
-
Adding ref
#775to changes→ <<cset e3c46f6bf88c>>
-
Correcting comment - web was shortened to we. Refs
#775→ <<cset d1f080b4c2f5>>
-
- changed status to resolved
Merged in issue775PHEXrayExportLRK (pull request #324)
Fixes
#775Approved-by: David Platten
→ <<cset b19d0bf7d640>>
- Log in to comment
Now writing cGycm2 values as well as units. Corrected number of projections in docs. Refs
#775→ <<cset 459f563eb74d>>