mAs charts are displaying microAs
Issue #901
resolved
Affects mammography scatter plot of mAs vs compressed breast thickness.
Radiographic charts are correct.
Comments (5)
-
reporter -
reporter Updating test file to reflect the correct mAs values for the mammo test data. Refs issue
#901→ <<cset 1a30c0f418f9>>
-
Correcting unit in doc string for RDSR version of _exposure to indicate uA.s (µA.s). Refs
#901→ <<cset 2aa1c5f27e1f>>
-
reporter Resolving merge conflict in changes.rst file. Refs issue
#901→ <<cset b1293bbcf1d6>>
-
reporter - changed status to resolved
Merged in issue901wrongMammomAsunits (pull request #450)
Issue901wrongMammomAsunits
Approved-by: Ed McDonagh
Fixes issue
#901→ <<cset 06894b496921>>
- Log in to comment
Added the optional data_point_value_multipliers argument to the main mammography dataframe creation routine. This allows a scaling factor of 0.001 to be applied to the uAs values to transform them into mAs. A scaling factor is required for every value_field if this argument is being used, so added scaling factors of 1 for the other value_names that could be present. Refs issue
#901→ <<cset f5ecb7e4c893>>