Added test to check that reference point dose is correctly imported from a Canon Ultimax-i where the values are stored in the RDSR in mGy. I can't run tests locally, so not sure if I have the correct event index at the moment. References issue #808
Correcting values in the new tests. Added test for total dose at RP. References issue #808
My colleague@Debbie Harries helped with this coding.
Added method to check dose at RP units and adjust if mGy found. References issue #808
Added tests for the new RP units check; also updated changes files. References issue #808
Added tblib to the tox file to invetsigate why tests are failing. References issue #808
Moved the section that tblib is in... References issue #808
Removed reference point dose units check in the TID 10003 section as this should be in mGy. Left the check in-place for reference point doses in TID 10003B (http://dicom.nema.org/medical/Dicom/2017e/output/chtml/part16/sect_TID_10003b.html), 10004 (http://dicom.nema.org/medical/Dicom/2017e/output/chtml/part16/sect_TID_10004.html) and 10007 (http://dicom.nema.org/medical/Dicom/2017e/output/chtml/part16/sect_TID_10007.html). References issue #808
Added test to check that reference point dose is correctly imported from a Canon Ultimax-i where the values are stored in the RDSR in mGy. I can't run tests locally, so not sure if I have the correct event index at the moment. References issue #808
Correcting values in the new tests. Added test for total dose at RP. References issue #808
My colleague@Debbie Harries helped with this coding.