Synergy Safety doorless Beam Terminate Elekta test description is incorrect

Issue #122 resolved
Former user created an issue

The test description is missing a step. You have to reset the key on the console before pressing the Reset Motors button in the tmt room.

It also says to finish running the 10MV 200MU beam. It should say the 6MV 200MU beam.

Comments (3)

  1. Log in to comment