Synergy Safety doorless Beam Terminate Elekta test description is incorrect
Issue #122
resolved
The test description is missing a step. You have to reset the key on the console before pressing the Reset Motors button in the tmt room.
It also says to finish running the 10MV 200MU beam. It should say the 6MV 200MU beam.
Comments (3)
-
repo owner -
repo owner Added additional step and changed 10MV to 6MV.
-
repo owner - changed status to resolved
- Log in to comment