Synergy Safety Doorless LPO System Functional test description is incorrect

Issue #123 resolved
Former user created an issue

The test says to reposition the Checkmate in the treatment beam by rotating the table column. We do not rotate the column on U21, 22 or 23 due to bed values. We move the bed longitudinally to place the checkmate. I think you could just say reposition the Checkmate in the treatment beam. That is enough.