Process Siemens Arcadis C-Arm dose reports
Add import routines to recognise and import data from Siemens Arcadis C-Arms with software version VC10A.
DICOM conformance statement is here: Arcadis VC10A
Uses object type EnhancedSRStorage
and can be identified by having a container with the following:
- CodeValue:
C-10
- CodingSchemeDesignator:
99SMS_RADSUM
- CodeMeaning:
Radiation Summary Report
The report is at a high level, but useful nonetheless. It contains the following according to the conformance statement:
- Cumulative fluoro time in seconds
- Cumulative dose area product in Gy·cm²
- DateTime of the start and end of the examination
- Cumulative air kerma in Gy, if configured.
In addition, the example of this report sent to me by Rachid Boucenna also includes a sequence for the total number of exposures.