Some Requested Procedure Code Sequences are longer than allowed 16 characters
Issue #501
resolved
And thus fail to import.
The standard dictates the VR of the CodeValue is SH, which is limited to 16 characters. It seems that Siemens permits longer strings to be used.
Propose changing the datatype in the database to relax this constraint.
Comments (3)
-
reporter -
reporter - changed status to resolved
Tested change in CodeValue field type with postgres, works fine. Requires migration. Fixes
#501→ <<cset a24382003940>>
-
reporter Changing code_value field types for long procedure codes for Stavanger. Refs
#501→ <<cset e614f831b02d>>
- Log in to comment
Changed code_value fields from SH/16 char fields to TextFields to ref
#501. Tested with SQLite3, need to test with PostgreSQL→ <<cset 2a2de5754fe0>>