Fails to import Toshiba Cath Lab RDSRs
Issue #227
resolved
Initial error is due to the RDSR having patient orientation information that is incorrectly coded in the extraction routine. The incorrect database fields were fixed in the last release.
Comments (4)
-
reporter -
reporter Patient Orientation Modifier should be at the same level as Patient Orientation. Corrected, and now seems to work. Need to check that the Siemens ones still work, and verify the data actually is in the database. Refs
#227→ <<cset 95bbe447383a>>
-
reporter - changed status to resolved
Fixed and tested.
-
reporter Added refs
#226,#227to the changes. Included information about Toshiba RDSRs in the release notes.→ <<cset 57a62944a83e>>
- Log in to comment
Changed Patient Orientation, Patient Orientation Modifier and Patient Table Relationship to use the newer cid fields. Added in some TODO information regarding where the Projection Eponymous Name field should be, both in RDSR and in models. Refs #55,
#227→ <<cset 039159c662ce>>